When you click on links to various merchants on this site and make a purchase, this can result in this site earning a commission. Affiliate programs and affiliations include, but are not limited to, the eBay Partner Network.
TS1 Touch screenPatient Monitor 6 Parameters
Introduction
The monitor is used to monitor, display, check, store, alarm and transmit many physiological parameters, including ECG (3/5/12 lead,arrhythmia analysis, ST segment analysis and HR ), RESP, SpO2, PR, NIBP, TEMP and IBP, etc. It is applicable for use by adult,
children and neonate.
The monitor can be used in medical institutions or during transport monitoring inside/outside the hospital environment, it is used by
clinical professionals or under their guidance or professionals via training, it is not suitable for use in family.
Function
1)5" touch screen, easy and convenient to operate, automatic lock screen function prevents misoperation.
2)Portable design (weight < 1.0 kg), flexible to transport.
3)Multi - interface selection, convenient to observe patient’s data trend.
4)Degree of protection provided by enclosure complies with IEC/EN 60529, waterproof degree: IP44
5)Fanless design, quiet, energy-saving and clean, which reduces the possibility of cross-infection.
6)Built-in large-capacity battery, meets the requirements of long-term operation.
Performance
(1)ECG
Lead type:
3-lead: I, II, III
5-lead: I, II, III, aVR, aVL, aVF, V
12-lead: I, II, III, aVR, aVL, aVF, V1,V2,V3,V4,V5,V6
Waveform
3-lead: 1 channel
5-lead: 2-channel
10-lead: 12-channel
Gain: 2.5 mm/mV, 5.0 mm/mV, 10 mm/mV, 20 mm/mV
Scan speed: 12.5 mm/s, 25 mm/s, 50 mm/s
HR
Measurement and alarm range:
Adult: 15 ~ 300 bpm
Pediatric/neonate: 15 ~ 350 bpm
Resolution: 1 bpm
Accuracy: ± 1 % or ± 1 bpm, whichever is greater
Alarm error: ± 2 bpm
ST-segment detection
Measurement and alarm range: - 2.0 mV ~ + 2.0 mV
Accuracy: - 0.8 mV ~ + 0.8 mV: ± 0.04 mV or ± 10 %, whichever is greater
Other range: unspecified
Arrhythmia analysis: 13 kinds
Pacemaker: yes
(2)RESP
Measurement mode: R-F (RA-LL) impedance
Respiration rate:
Measurement range: 0 ~ 150 rpm
Resolution: 1 rpm
Accuracy: 7 ~ 150 rpm, ± 2 rpm or ± 2 %, whichever is greater;
0 ~ 6 rpm, unspecified
Apnea alarm: 10 ~ 40 s
Scan speed: 12.5 mm/s, 25 mm/s
(3)NIBP
Measurement mode: Oscillometry
Working mode: Manual/AUTO/Continuous
Measurement interval in AUTO mode: 1 / 2 / 3/ 4 / 5 / 10 / 15 / 30 / 60 / 90 / 120 / 240 /480 /960 minutes
Measurement period in Continuous mode: 5 minutes
Measurement and alarm range: 10 ~ 270 mmHg
Resolution: 1 mmHg
Cuff pressure accuracy: ± 3 mmHg
Measurement accuracy:
Maximal mean deviation: ± 5 mmHg
Maximal standard deviation: 8 mmHg
Over-pressure protection:
Adult mode: 297 ± 3 mmHg
Pediatric mode: 240 ± 3 mmHg
Neonatal mode: 147 ± 3 mmHg
(4)SpO2
Measurement and alarm range: 0 ~ 100%
Resolution: 1 %
Measurement accuracy: 70 % ~ 100 %: ± 2 %;
0 % ~ 69 %: unspecified
(5)PR
Measurement and alarm range: 25 ~ 250 bpm
Accuracy: ± 2 bpm or ± 2 %, whichever is greater
(6)TEMP
Channel: dual-channel
Measurement and alarm range: 0 ~ 50 ℃
Resolution: 0.1 ℃
Accuracy: ± 0.1 ℃
Accessories
CategoryAccessory name
Main unitTS1 main unit
ECG cable5-lead, American standard, TPU
SpO2 probe (integrated)Digital fingertip SpO2 probe for adult (CMS-N-SPO2 6P, 3M, yellow)
TEMP probeR25=2.252K TEMP probe, skin surface type, CMS-N-TEMP 2P, PVC, L=3M
NIBP cuffBP extension tube, L=3M
Adult cuff (25 ~ 35 cm)
Physical characteristic
Dimension: 142 mm × 67 mm × 102 mm
Weight: < 1 kg